Dimensions (W x D x H)
76 x 81 x 130 cm (30 x 32 x 51 in.)
86 kg (190 lb)
Power requirements
(includes compressor)
115 VAC, 60 Hz, 4.2 A or 6.4 A with controller
compartment heater
203 VAC, 50 Hz, 2.7 A or 4.1 A with controller
Overload protection
115 VAC: 7.5 A circuit breaker
230 VAC: 5.0 A circuit breaker
Compressor
1/5 HP
115 VAC: 115℃ (239℉) thermal overload protector,
7.1 locked rotor amps
230 VAC: 120℃ (248℉) thermal overload protector,
7.6 A peak start current
Operating temperature
0 to 50℃ (32 to 122℉); with AC battery backup:
0 to 40℃ (32 to 104℉); with controller compartment
heater: -40 to 50℃ (-40 to 122℉); with controller
compartment heater and AC battery backup: -15 to 40℃
(5 to 104℉)
Storage temperature
-30 to 60℃ (-22 to 140℉)
Relative humidity
0 to 95%
Installation category,
pollution degree
II, 2
Protection class
l
Temperature control
4 (±0.8)℃ (39(±1.5)℉)
[Radio frequency interference in the 30 to 50 MHz range
can cause a maximum temperature change of 1.3℃ (34.3℉). Adjust the set point temperature to 2 to 10℃ (35.6 to 50℉) to correct this interference.]
Enclosure
IP24, low-density polyethylene with UV inhibitor
Sample bottle capacity
Single bottle: 10 L (2.5 gal) glass or polyethylene, or 21 L (5.5 gal) polyethylene
Multiple bottles: two 10 L (2.5 gal) polyethylene and/or glass, four 10 L (2.5 gal) polyethylene and/or glass, eight 2.3 L (0.6 gal) polyethylene and/or 1.9 L (0.5 gal) glass, twelve 2 L (0.5 gal) polyethylene (AWRS only), twenty-four 1 L (0.3 gal) polyethylene and/or 350 mL (12 oz.) glass
Certifications
AC power supply: cETLus, CE